Specific Conductance: Theoretical Consideration and Application to Analytical Quality Control, 1986
Bibliographic details
Summary
This report considers several theoretical aspects and practical applications of specific conductance to the study of natural waters. A review of accepted measurements of conductivity of secondary standard 0.01 N KCI solution suggests that a widely used alogrithm for predicting the temperature variation in conductivity is in error. A new alorithm is derived and compared with accepted measurements. Instrumental temperature compensation circuits based on 0.01 N KCI or NaCl are likely to give erroneous results in unusual or special waters, such as seawater, acid mine waters, and acid rain. An approach for predicting the specific conductance of a water sample from the analytically determined major ion composition is described and critically evaluated. The model predicts the specific conductance to within 8 percent (one standard deviation) in waters with specific conductances of 0 to 600 uS/cm. Application of this appraoch to analytical quality control is discussed.